3D microdosimeters are designed to measure the effects of ionizing radiation at the scale of biological cells. That level of measurement matters in applications where radiation quality can vary over short distances, including charged-particle therapy and radiation protection research. Work reported by the Australian Institute of Nuclear Science and Engineering (AINSE) describes silicon and diamond detector structures developed and evaluated for this purpose, including a silicon-on-insulator (SOI) device tested in a carbon-ion beam.

The engineering challenge is not simply to detect radiation. A useful microdosimeter must collect charge from a small sensitive volume and provide information about how energy is deposited in that volume. Three-dimensional detector designs, ion-beam testing and fine spatial resolution offer ways to investigate these requirements.

Why 3D microdosimeters matter

Microdosimetry examines the effects of ionizing radiation at biological cell scales. It can help researchers monitor, quantify and model radiation damage in mixed radiation environments. The source identifies potential applications in cancer treatments using protons and heavier ions, radiation safety in areas such as aviation and mining, and space exploration.

In particle therapy research, measurements need to capture changes in radiation effects across a treatment field. The reported work focused in part on the spread-out Bragg peak (SOBP) of a carbon-ion therapeutic beam. This region is relevant to understanding dose delivery, while the distal part of the peak and the fragment tail present measurement challenges when spatial resolution is limited.

A key objective was to assess whether solid-state detector structures could provide microdosimetric information at a finer scale. The research examined silicon detectors with three-dimensional electrode arrangements, an SOI microdosimeter, and a prototype diamond detector with separated sensitive regions.

3D detector structures and charge collection

In a 3D detector, electrodes and active edges are formed within the detector material using micromachining techniques. This arrangement differs from a structure in which the relevant electrodes are placed only on the material’s surface. The research investigated how these internal geometries affected charge collection and whether they could support low-noise operation.

Researchers used ion beam induced charge (IBIC) measurements at ANSTO to examine detector responses. In this technique, an ion beam is directed at a device and the resulting charge response is measured. The reported studies included 3D silicon detectors with columnar electrodes and tests of their response under different applied biases. Several geometries showed full depletion under low applied bias, an observation relevant to the development of low-noise detectors. The work also investigated how fast-neutron and gamma radiation affected charge-collection efficiency.

The project then applied 3D fabrication methods to an SOI microdosimeter design. A large-area, ultra-thin detector was fabricated on a high-resistivity silicon layer. Its supporting silicon wafer was removed beneath the active area, while the silicon dioxide layer remained. The resulting structure was evaluated for microdosimetric measurements in a carbon-ion beam.

Measuring carbon-ion radiation at fine scale

Using the ultra-thin detector, the research team obtained microdosimetric spectra along the SOBP of a 290 MeV/u carbon-ion beam. Those measurements were used to derive relative biological effectiveness (RBE) values with sub-millimetre spatial resolution. RBE is a way of comparing the biological effect of one type of radiation with a reference; in this work, the reported RBE10 values were derived from the microdosimetric spectra using a model.

The detector-derived results were compared with RBE10 values based on measurements from a tissue-equivalent proportional counter (TEPC). The reported values matched well in both the plateau and SOBP regions. In the distal part of the SOBP and the fragment tail, the detector-derived values showed a better match to theoretical simulations than the TEPC-derived values. The report notes that TEPC measurements in these regions can be affected by volume averaging.

These results demonstrate the potential value of a solid-state microdosimeter when measurements must resolve changes over short distances. They are research findings, rather than a general performance guarantee for all detector configurations or treatment settings.

Diamond detectors and pixel separation

The project also explored laser-milled trenches as a way to create separated sensitive regions in diamond detectors. IBIC measurements showed that the trenches affected the lateral confinement of charge collection, effectively forming a pixelated detector structure. The prototype contained four pixel structures separated by the milled trenches.

This work points to another possible route for building microdosimeters. Silicon and diamond each featured in the reported research, but the article’s source does not establish that one material is universally preferable. Instead, the studies show how detector geometry and fabrication methods can be investigated through charge-collection measurements.

Practical implications for detector development

For engineers and technical teams working on radiation instrumentation, the reported research highlights several considerations:

  • Geometry affects measurement: Electrode arrangement and the definition of sensitive volumes are central to charge collection.
  • Spatial resolution matters: Fine-scale measurements can reveal variations that may be obscured by volume averaging.
  • Testing must match the application: IBIC provides a means to characterize detector response, while beam measurements can assess performance in a relevant radiation field.
  • Fabrication is part of the design: Micromachining, wafer thinning and laser-milled trenches were all used to create or modify detector structures.

The work also illustrates the role of collaboration among detector researchers, radiation facilities and medical-physics teams. The SOI device was tested in collaboration with the National Institute of Radiological Sciences in Japan, while fabrication and characterization drew on specialist expertise across several institutions.

Conclusion

3D microdosimeters offer a research pathway toward measuring radiation effects with sensitive volumes and spatial detail suited to cell-scale questions. The reported silicon and diamond studies show how electrode geometry, material processing and charge-collection characterization can contribute to that goal. For organizations developing radiation detectors, the practical lesson is to evaluate the complete measurement chain: detector structure, fabrication, response testing and the spatial detail required by the intended application. Continued development of these approaches may support research in charged-particle therapy and radiation protection, while application-specific validation remains essential.