Continuous, autonomous monitoring places demanding requirements on radiation detection instrumentation: systems must identify relevant signals, manage background events and operate reliably without unnecessarily complex hardware. A research prototype radioxenon detection system explores one way to address those requirements by combining coplanar cadmium zinc telluride (CZT), a silicon photomultiplier (SiPM) array and a plastic scintillator in a compact beta-gamma coincidence detector.

The work is relevant beyond nuclear monitoring as an example of how detector materials, signal conditioning and real-time digital processing can be brought together in a small instrument. Its results are preliminary, however: the prototype demonstrates coincidence measurements and characterizes selected energy peaks, but does not yet establish a minimum detectable concentration for atmospheric monitoring.

Why beta-gamma coincidence matters

Radioxenon isotopes are monitored because their characteristic decay emissions can help distinguish releases associated with different nuclear processes. Several isotopes emit combinations of beta particles or conversion electrons alongside X-rays or gamma rays. Detecting these emissions in coincidence—within a defined time window—provides a way to identify relevant decay signatures while rejecting events that do not match the expected pattern.

Energy resolution is important to this process. A detector’s measured peaks define regions of interest (ROIs) on a coincidence plot. Narrower, clearly separated peaks can support smaller ROIs, reducing overlap between signatures and limiting the chance that random background coincidences fall within a region. The 30 keV X-ray is particularly important because it serves as a coincidence identifier for several radioxenon isotopes.

How the radioxenon detection system is assembled

The prototype brings together two detection channels. A coplanar CZT crystal detects X-rays and gamma rays, while a cylindrical plastic scintillator serves both as the beta-detection medium and as a gas cell for the radioxenon sample. An array of SiPMs reads the scintillator’s light output. The detector components and analog front-end electronics are mounted on a custom printed circuit board (PCB).

CZT for room-temperature photon detection

CZT provides compact gamma-ray spectroscopy without requiring the detector to be cooled with liquid nitrogen. In a coplanar-grid design, signals from collecting and non-collecting anodes are combined by subtraction. This approach makes the measured signal less dependent on where an interaction occurs within the crystal, while keeping the readout to two anode signals. In the prototype, the crystal measures 1 × 1 × 1 cm³.

Plastic scintillator and SiPM light readout

The plastic scintillator is shaped as a small, cylindrical gas cell. Its walls are intended to stop beta particles and conversion electrons while allowing low-energy X-rays and gamma rays to pass through with limited attenuation. The geometry also provides broad coverage for charged particles. Light from the scintillator is collected by a 3 × 3 SiPM array coupled to its base.

Compared with a photomultiplier tube (PMT), SiPMs are small and operate at lower voltage. Their compact form supports mounting the optical sensors alongside other detector components on the PCB. This integration is useful where size and straightforward assembly matter, although system performance still depends on the complete optical, electrical and mechanical design.

Real-time coincidence processing

Signals from the CZT and SiPM channels pass through analog preamplification and into a two-channel digital pulse processor. Its 200 MHz, 12-bit analog-to-digital converter samples the detector pulses, and an onboard field-programmable gate array (FPGA) applies the coincidence logic.

For an event to be accepted, both channels must exceed their trigger thresholds and fall within the configured coincidence timing window. Events that fail either condition are rejected; accepted signals are sent to a computer for display and analysis. This arrangement moves coincidence discrimination into the digital processing chain rather than relying on separate instruments to identify paired events.

Timing and filtering settings were adjusted during testing. The coincidence window was initially set to 0.8 microseconds and later changed to 0.75 microseconds based on measurements with radioxenon samples. A trapezoidal digital filter was used to reduce unwanted noise while allowing charge collection. These are experimental settings for this prototype, not universal values for other detectors.

What the measurements show—and what they do not

Initial characterization used a cesium-137 source, followed by samples containing xenon-135 and xenon-133/xenon-133m. The prototype recorded the expected types of coincident signals, including the 250 keV gamma-ray signature associated with xenon-135 and the 30 keV and 81 keV features associated with xenon-133 measurements. The reported energy resolution for the 30 keV peak was 30.0%; the 81 keV peak was 12.8%. The 250 keV result was reported as 5.33% in the results discussion.

These measurements support the proof of concept: the two detector channels and FPGA-based processing can identify beta-gamma coincidences. They also show why low-energy resolution remains a key design consideration. Because ROI size affects how coincidence events are separated from background, the measured peaks must be refined and evaluated under fuller operating conditions before detection sensitivity can be assessed.

The researchers identify several next steps: measure background, repeat isotope measurements after further optimization, assess xenon-131m, and calculate minimum detectable concentrations. They also describe potential future changes, including adding or enlarging CZT crystals to increase photon-sensitive area. These are planned investigations, not demonstrated performance improvements.

Practical implications for instrument designers

For engineers and technical decision-makers, the prototype illustrates the value of treating detector geometry, sensor selection, analog electronics and digital processing as one system. A compact PCB-based arrangement can simplify physical integration, while real-time FPGA coincidence logic can make event selection part of the acquisition architecture. At the same time, the gas-cell design, light collection, signal noise, energy calibration and timing configuration all influence whether the resulting data are useful.

The practical takeaway is to distinguish a promising architecture from a field-qualified instrument. This radioxenon detection system demonstrates compact, room-temperature beta-gamma coincidence measurements using CZT, SiPMs and plastic scintillation. Establishing operational sensitivity will require the planned background characterization, optimized measurements and minimum detectable concentration analysis. For instrumentation teams, that progression—from integrated prototype to quantified performance—is essential when assessing whether a design is ready for autonomous monitoring.