Ear canal movement measurement offers a potential route to estimating bite force with a wearable sensor positioned in the ear. In a small foundational study, researchers compared an earphone-type optical sensor with measurements of masseter muscle activity and occlusal force. The work explores a sensing approach that could avoid placing a force sensor in the mouth during eating, while also highlighting the calibration and validation challenges that arise when measurements depend on individual anatomy.
How ear canal movement measurement works
The approach relies on the mechanical relationship between jaw activity and the ear canal. During occlusion, the shape of the ear canal can change because of nearby masticatory muscles and the temporomandibular joint. An earphone-style sensor measures these changes optically rather than using a sensor inserted between the teeth.
The prototype used an infrared LED and a phototransistor. The LED illuminated the ear canal, and the phototransistor detected reflected light. Changes in reflected light altered the sensor output, providing a signal associated with changes in ear canal shape. The sensor was housed in a modified commercial earplug.
This is an indirect measurement: the device does not sense bite force at the teeth. Instead, it records a physiological movement associated with jaw loading, then uses a relationship between the sensor signal and a reference force measurement to estimate occlusal force.
Testing the relationship between movement and force
Five healthy adults, aged 21 to 26, each completed six measurement runs. During each run, they gradually increased pressure while biting the sensing area of an occlusal force meter with a second molar, then relaxed the pressure. Researchers simultaneously recorded the ear sensor output, surface electromyography (EMG) from the masseter muscle, and force-meter output.
Signals were sampled at 100 Hz with 12-bit resolution. The experiment extended earlier work on light chewing, which examined forces of approximately 40 N, by investigating stronger bites exceeding 400 N. The study therefore tested whether the ear sensor signal remained related to force over a broader loading range.
The researchers used Pearson correlation to assess how pairs of signals varied together. They also calculated partial correlations to examine the relationship between ear sensor output and occlusal force while accounting for EMG, and vice versa. This distinction matters because both ear canal movement and muscle electrical activity may change as bite force changes. A partial correlation can help assess whether the sensor-force relationship is present beyond their shared association with muscle activity.
What the results indicate
The absolute average correlation between ear sensor output and occlusal force was at least 0.9514 for each participant. The direction of the relationship differed: it was positive for two participants and negative for three. The authors attributed this variation as a possibility to individual differences in ear canal shape, which can affect how reflected infrared light reaches the phototransistor.
After accounting for EMG, the absolute average partial correlation between ear sensor output and occlusal force ranged from 0.6161 to 0.8286. This suggests that the relationship was not explained entirely by the measured masseter activity in these trials. It does not, however, establish that the ear sensor measures force independently of all other influences.
For force estimation, the researchers applied single regression analysis separately for each participant. Five runs formed the training data, and the remaining run served as the test data; this process was repeated across all six runs. The reported average root-mean-square error (RMSE) values ranged from 0.0338 to 0.0969. These errors were calculated using the study’s converted signal values, so they should not be interpreted directly as errors in newtons.
Engineering considerations and limitations
The findings demonstrate a promising signal relationship under controlled conditions, not a ready-to-deploy measurement system. The study included only five young, healthy participants, and the regression approach was individualized. Differences in ear canal geometry may change both the magnitude and direction of the sensor response, making calibration an important design consideration.
Fit and motion are additional concerns. The ear sensor was housed in an earplug that did not necessarily fit every participant perfectly. Head movement could shift the sensor and introduce motion artifacts, potentially changing the optical measurement independently of jaw loading. The experiment asked participants to keep their heads still, so performance during unrestricted movement remains an open question.
The researchers also noted that the sensor response might eventually saturate outside the tested force range. Anatomical limits on ear canal movement could restrict how the signal changes at higher loads. Further work would need to assess robustness across more participants and conditions, improve the estimation and signal-processing methods, and determine appropriate device specifications through additional evaluation.
Practical implications for wearable sensing
For engineers developing wearable measurement systems, the study illustrates how a noninvasive sensor can infer a target variable from a related physical signal. It also shows why correlation alone is not enough: calibration, subject-to-subject variation, movement artifacts, operating range, and validation on held-out data all affect whether an estimate is useful.
The proposed earphone-type device could ultimately support occlusal force monitoring during eating without an intraoral force sensor or jaw-mounted EMG electrodes. In this study, however, that application remains a development goal rather than a demonstrated clinical or commercial capability. The practical takeaway is that ear canal movement is a measurable candidate signal for force estimation, but reliable use will depend on better characterization of individual fit, motion sensitivity, and estimation accuracy.