Positron burst detection presents a measurement challenge: a short pulse can contain many positrons, while conventional detectors may cover only a limited area and experience signal pile-up. A pixelated approach offers another way to collect annihilation timing information. The system described here combines calcium-doped LYSO scintillators, silicon photomultiplier (SiPM) arrays and DRS4 waveform digitizers to record signals across a large number of channels.

The design is intended for positron annihilation lifetime spectroscopy (PALS), particularly measurements using intense, short-pulsed positron beams. Its central idea is to distribute detectors spatially and use the beam’s generation signal as the timing start, rather than relying on a conventional two-detector coincidence arrangement for each event.

Why positron burst detection needs a different approach

In PALS, the time between positron formation and annihilation provides information used to study material microstructure. A short-pulsed beam concentrates positrons into a brief burst. Depending on the source, bursts may contain from 105 to 108 positrons, creating the potential to observe changes that develop over time in materials.

That intensity also puts pressure on the detection system. Traditional arrangements commonly use photomultiplier tubes and bulk scintillators. Two detectors can capture only a limited solid angle, and high event rates can cause signals to pile up or saturate. These limitations make it difficult to use all the annihilation events in an intense burst.

How the spatial timing method works

The proposed design places pixelated detectors over a hemispherical surface, covering a 2π solid angle. Because annihilation gamma photons are emitted in pairs and in isotropic directions, photons from the events can reach detectors across the surface. The burst-generation signal serves as the start time; each detector channel’s signal supplies a stop time. The resulting distribution of start-to-stop times forms the lifetime spectrum.

This is a time-to-space measurement concept: instead of depending on a small number of detector pairs, it uses many independently read channels to collect timing events across the array. The approach is intended to increase use of the positrons in a burst. In the prototype described, the channel count is on the order of 103; acquiring a spectrum with about 106 events would therefore require roughly 103 bursts.

Timing performance depends on both the width of the positron pulse and the timing resolution of the detectors. The array architecture does not remove that constraint; it provides a way to gather events spatially while preserving the timing information needed for PALS.

Detector materials and readout choices

LYSO balances light collection and stopping power

The scintillator must convert gamma-ray interactions into enough light for accurate timing while also efficiently stopping 511 keV photons. The design uses calcium-doped lutetium yttrium oxyorthosilicate (Ca:LYSO). In simulations comparing several fast inorganic scintillators with a fixed 3 × 3 mm² cross-section, a 5 mm LYSO crystal produced a full-energy detection ratio comparable to longer crystals of some alternatives. The comparison considered materials including GAGG, BaF2 and LaBr3.

Light collection is another important factor. For a scintillator-based detector, timing performance is affected by scintillation rise and decay times and the amount of light collected. More collected light can support better timing when the scintillator’s time characteristics are fixed. LYSO’s natural background radiation from lutetium-176 was noted, but the authors considered it negligible within the system’s very short signal timing window.

SiPM arrays and waveform digitization

Each detector couples a 3 × 3 × 5 mm³ LYSO crystal to a 3 × 3 mm² SiPM. The documented system has 32 modules, each containing an 8 × 8 crystal assembly and a 64-channel SiPM array, for a total of 2,048 detector channels. The modules are distributed across a 40 cm diameter hemisphere.

SiPMs provide the photosensitive element, while DRS4-based electronics capture the analog waveforms. The DRS4 is a switched-capacitor sampling chip; its stored waveform samples can be read out for external digitization. An FPGA then applies digital constant fraction discrimination to determine signal timing. This separates waveform acquisition from the subsequent digital timing decision.

Before testing, the single-channel detectors were optimized through changes to scintillation-light collection, SiPM readout electronics and temperature stability. The digital electronics were also checked using a shared pulse sent to two channels. After DC-offset calibration, the reported electronics time jitter was 21.5 ps.

What the preliminary timing result shows

Coincidence timing was evaluated with two identical single-channel detectors positioned 5 cm apart, with a 22Na point source between them. The test selected double 511 keV coincidence events. With 5 mm crystals, the best reported coincidence timing resolution was 84 ps. Longer crystals gave different results, with the 10 mm and 15 mm configurations showing wider timing values in the reported measurements.

This result is a detector-level coincidence measurement, not a demonstration of a complete positron-burst lifetime spectrum from the full 2,048-channel array. The work describes module- and system-level studies as ongoing. It also notes that clock accuracy and channel non-uniformity may affect the timing performance of a multichannel DRS4 board.

Practical implications for measurement-system design

For engineers and technical teams working on specialized instrumentation, the design illustrates how detector geometry, sensor selection and digital signal processing must be considered together. A large channel count can support spatial event collection, but it also makes consistent waveform acquisition and timing calibration important. The reported 84 ps result provides a preliminary reference for two single-channel detectors under the stated test conditions; it should not be treated as a performance specification for the complete array.

Overall, this positron burst detection concept addresses the limits of conventional arrangements by spreading detectors across a hemisphere and recording timing signals channel by channel. Its combination of Ca:LYSO crystals, SiPM arrays and DRS4 waveform digitization offers a framework for high-intensity pulsed PALS measurements. Further system-level validation is needed to establish how the full array performs with positron bursts and how reliably it captures dynamic changes in materials.