Printed OLEDs bring together organic semiconductor materials and scalable fabrication methods to create light sources that can be thin, flexible and tailored to specific wavelengths. In biomedical devices, these characteristics could support optical sensors that conform to the body more closely than rigid components. Research on OLED-based pulse oximeters illustrates both the promise of this approach and the manufacturing challenges that remain.
Why printed OLEDs are relevant to wearable sensors
Organic light-emitting diodes use organic semiconductor materials rather than the inorganic semiconductors found in conventional LEDs. Their emission characteristics can be adjusted through material selection, while solution-based fabrication methods—including spin coating and blade coating—can form light-emitting layers on substrates such as plastic.
This combination is relevant to wearable optoelectronics because it offers a path to devices that are lightweight, flexible and scalable in area. Biomedical applications place particular demands on the light source: its emission spectrum and the amount of light reaching tissue must suit the measurement. For pulse oximetry, for example, the sensor uses light at different wavelengths to detect changes associated with blood volume and oxygenation.
From OLED prototypes to pulse oximetry
A pulse oximeter measures optical changes in tissue. As blood volume in the arteries changes with each heartbeat, the amount of light reaching a photodetector varies. This changing signal is called a photoplethysmogram, or PPG. Heart rate can be derived from the pulse pattern, while oxygen saturation is calculated using signals measured at different wavelengths.
In a proof-of-concept transmission-mode device, red and green polymer OLEDs were paired with photodetectors. Although clinical pulse oximeters commonly use red and near-infrared light, green was used in this demonstration because solution-processable visible emitters were more developed. The OLEDs produced a clear PPG, and the all-organic optoelectronic sensor measured heart rate and oxygenation within 2% of a commercially available hospital-grade pulse oximeter.
This result demonstrated that organic light sources could operate in a biomedical sensing system—not that every component or manufacturing step was already ready for high-volume production. Subsequent work addressed the electrode, coating process and emission spectrum, all of which affect how a device might be made and used.
Addressing flexible-device manufacturing constraints
Replacing brittle transparent electrodes
Indium tin oxide (ITO) is commonly used as a transparent anode in OLEDs, but its brittleness makes it a less suitable choice for a device intended to flex with the body. One alternative explored was a tungsten trioxide/silver/tungsten trioxide (WO3/Ag/WO3) electrode on flexible polyethylene naphthalate (PEN). The polymer layers were blade coated, a method intended to improve throughput compared with spin coating.
The demonstrated flexible near-infrared OLED used a thermally evaporated WO3/Ag/WO3 anode. It emitted at a peak wavelength of 735 nm and achieved 2.17% external quantum efficiency at a current density of 100 mA/cm². These results establish a device-level demonstration; they do not remove the need to assess electrode processing, device integration and production requirements for a specific application.
Moving more of the structure to printing
For a more scalable manufacturing route, the research also examined printing the electrode layers. WO3 layers were blade coated from a nanoparticle ink, while the silver layer was still thermally evaporated. Red OLEDs made on these partially printed electrodes emitted at 617 nm and produced a forward flux of 1–10 mW/cm² per pixel area, a range described as compatible with optical biomedical-device wavelength and irradiance requirements.
Printing the silver layer remained a challenge. Further development of the silver ink was needed to achieve a fully printed WO3/Ag/WO3 electrode. This distinction matters for process planning: demonstrating printed polymer layers or printed portions of an electrode is not the same as demonstrating a fully printed device structure.
Why near-infrared emission matters
For optical measurements, emission wavelength affects how light interacts with tissue. The work explored shifting an OLED’s output deeper into the near-infrared by increasing the emissive-layer thickness from 150 to 250 nm. This changed optical interference within the device, producing a secondary emission peak at 815 nm alongside the principal peak near 745 nm.
In a PPG comparison, the calibrated signal from the OLED with the 815 nm secondary peak was 32.87% stronger than the signal from the 745 nm OLED. The stronger signal was recorded at a lower irradiance: 0.33 mW/cm² compared with a maximum of 1 mW/cm² for the 745 nm device. The comparison highlights that selecting an emission spectrum is not just a materials question; it can affect the optical signal available to the sensor.
Designing for the body as well as the process
Flexible OLED-based oximeter prototypes were also assessed with physician input on potential use in areas including ambulatory emergency care, chronic respiratory disease management, intraoperative monitoring and pediatrics. This feedback informed how and where a sensor could interface with the body. PPG signals were acquired from flexible sensors placed on the abdomen and back—locations beyond the conventional finger-probe arrangement.
For engineers and integrators, this points to a broader design task: the light source, detector, substrate, electrode stack and attachment method must work as a system. A flexible emitter alone does not determine whether a sensor will suit a particular body location or monitoring scenario.
Practical implications for development
Printed OLED research demonstrates a credible path toward flexible biomedical light sources, while also making the remaining process-development work clear. Key considerations include:
- Wavelength selection: Match the OLED emission to the optical measurement, and consider spectral width and secondary peaks.
- Electrode design: Evaluate alternatives to ITO when flexibility is important, while accounting for which layers still require evaporation.
- Coating method: Blade coating offers a route beyond spin-coated layers, but ink behavior and layer compatibility remain central to process development.
- Device integration: Assess the complete emitter–tissue–detector arrangement and the intended sensor placement.
Overall, printed OLEDs have demonstrated optical sensing potential in pulse oximetry and progress toward flexible, more print-compatible device structures. The practical opportunity for manufacturers and system designers lies in aligning material and process choices with the sensing requirements, while distinguishing successful prototypes from a fully printable production process.