Subnanoliter NMR brings magnetic resonance spectroscopy to samples too small for conventional inductive spectrometers to detect effectively. The approach combines a miniaturized CMOS detector with microfluidic structures that position and hold a sample close to an integrated planar coil. This research also demonstrates microscale electron paramagnetic resonance (EPR) and ferromagnetic resonance (FMR), illustrating how detector design and microfabrication can extend measurement capabilities to very small volumes and structures.

Why sample size challenges magnetic resonance

NMR and EPR provide information about chemical composition, molecular environments and magnetic behavior. But as sample volume shrinks, the number of contributing spins also falls, making the detected signal more difficult to distinguish. Conventional inductive instruments are designed for much larger samples, so their detectors are poorly matched to many cells, microorganisms and other volume-limited specimens.

Miniaturizing the detector helps address this mismatch. A small coil can improve the filling factor—the proportion of the detector’s sensitive region occupied by the sample—and increase the signal contribution from a nearby specimen. The CMOS NMR system described in the work uses an integrated planar microcoil for both excitation and detection, with a detector geometry optimized for volumes around 200 pL.

Position matters as much as volume. The coil’s sensitivity is not uniform: it decreases with distance from the coil. A sample positioned farther away therefore contributes less signal, potentially requiring longer measurements. The microfluidic structure must place the specimen accurately near the coil while maintaining a reliable seal and enabling controlled sample handling.

Microfluidics for subnanoliter NMR

Microfluidic channels provide a way to load, move and trap small liquid samples at the detector’s most sensitive region. In this work, channels were built using conventional microfabrication methods and two-photon polymerization, a form of high-resolution 3D printing with a reported lateral resolution of 300 nm.

These fabrication routes serve different design needs:

  • Conventional microfabrication: Suitable for production at larger scale and for channel designs that do not require complex three-dimensional geometry or resolution beyond photolithography.
  • Two-photon polymerization: Useful for rapidly prototyping intricate 3D structures, including devices designed to trap or culture biological specimens with differing sizes and fluidic behavior.

A thin separation between the sample and the integrated coil helps preserve sensitivity. The research reports using separation layers below 10 μm to achieve acceptable filling factors with a limited sensitivity penalty. This highlights a practical design tradeoff: the channel must provide enough structure for fluid handling and sealing, but excessive distance between the sample and coil reduces the detected signal.

Improving spectral quality through channel design

For NMR, microfluidic geometry affects more than sample placement. Dead volume—the portion of fluid that is not effectively measured—can waste scarce material. A channel designed around the active detection region can reduce this unused volume. The geometry can also help limit magnetic-field nonuniformities that broaden spectral features and reduce resolution.

Using 3D-printed microfluidic systems, the work reports spectral resolution down to 0.007 ppm for liquid samples of 100 pL. The flexibility of the printed design made it possible to tailor the channel to the small active volume, rather than adapting the sample to a larger, fixed measurement space.

The same approach enabled NMR measurements of intact biological specimens submerged in liquid media, including tardigrade ova and sections of adult Caenorhabditis elegans. The reported sensitivity was 2.5 × 1013 spins per square root hertz, sufficient to detect highly concentrated endogenous compounds in active volumes down to 100 pL, with measurement times as short as three hours.

Extending CMOS detection to EPR and FMR

The work also paired a CMOS EPR detector with commercially available capillaries to measure subnanoliter liquid samples and frozen solutions. Frozen samples are relevant where paramagnetic species have relaxation times too short for measurement at room temperature. The experiments provide a step toward studying such volume-limited biological samples, while demonstrating that miniature detectors can be used across different magnetic resonance methods.

For FMR, the detector was applied to magnetic microbeads and patterned magnetic materials. The bead-sensing method detects a change in magnetic susceptibility under FMR conditions. The reported variation was 20 times greater than the change in magnetization measured in static conditions by other approaches described in the work. The method demonstrated single-bead sensitivity across a sensing area of approximately 5 × 104 μm2.

FMR measurements were also demonstrated on individual patterned dots made from permalloy (Ni80Fe20) and yttrium iron garnet (YIG, Y3Fe5O12). These experiments show the detector’s use for investigating magnetic structures at micro- and submicrometer scales.

Practical implications for microscale instrumentation

For engineers developing analytical instruments, the central lesson is that detector sensitivity cannot be treated separately from sample handling. Coil dimensions, channel geometry, sample-to-coil spacing, sealing materials and fluidic interfaces all influence whether a small specimen can be positioned and measured effectively.

Fabrication method should follow the application. Established microfabrication processes are appropriate when the design is relatively planar and repeatable production is important. Two-photon polymerization offers design freedom for complex trapping structures and rapid prototyping. Neither route removes the need to optimize the interface between the microfluidic device and the detector.

Overall, subnanoliter NMR and related microscale EPR and FMR methods show how integrated CMOS coils can be combined with purpose-built microfluidics to examine very small liquid samples, biological entities and magnetic structures. The practical opportunity lies in designing the sensing element and sample-handling system together, so that scarce material is delivered precisely where the detector can use it most effectively.