Access Official Technical PDF

Share Summary:LinkedInEmail

Electrochemical gas sensors translate gas concentration into a small output current, often spanning tens of nanoamps to tens of microamps. For a monitoring system to use that signal, it must be converted into a voltage and filtered before reaching a comparator or analog-to-digital converter (ADC). A transimpedance amplifier (TIA) performs this conversion, while its feedback components set the signal gain and help limit noise bandwidth. In a gas-sensing application, careful selection of the amplifier and feedback network supports accurate measurement without requiring a high-bandwidth signal path.

How a transimpedance amplifier processes sensor current

A basic TIA uses an operational amplifier with a feedback resistor, RF, between its output and inverting input. The sensor current, IS, is converted to an output voltage, with the transimpedance gain principally set by RF. In practical terms, increasing the feedback resistance produces a larger voltage change for a given change in sensor current. The resistor must therefore be chosen with both measurement sensitivity and the amplifier’s available output range in mind.

A feedback capacitor, CF, works with RF to limit the circuit’s closed-loop bandwidth. This creates a low-pass response that reduces the range of frequencies passed to downstream electronics. The sensor itself can be represented in a simplified model as a current source in parallel with a sensor capacitance, CS. For the electrochemical sensor considered in the design, CS may be in the millifarad range. The sensor may also require a load resistor, RL, specified by its manufacturer; typical values described for this application range from 10 to 100 ohms.

CF has a second role beyond filtering: a minimum capacitance is needed to ensure TIA stability. The design therefore involves balancing noise filtering, the required signal bandwidth, and stable amplifier operation—not simply selecting a capacitor to obtain a preferred cutoff frequency.

Why gas sensing calls for low bandwidth and low power

Electrochemical gas sensors can be noisy, so filtering is useful before their signals enter a comparator or ADC. At the same time, these sensors respond slowly to changing gas concentrations, typically over a timescale of seconds. A fast signal path is not generally necessary for this type of measurement. A bandwidth greater than 1 Hz can be appropriate, providing a signal path faster than the sensor’s response while still limiting noise.

These conditions make a precision, low-bandwidth, nano-power operational amplifier a relevant choice. The example design uses a 3.3 V supply and the LPV821, a zero-drift amplifier. Its listed characteristics include 650 nA quiescent current, 1.5 µV input offset voltage, 20 nV/°C offset drift, 7 pA input bias current, and 8 kHz gain-bandwidth product. Low input offset matters because offset can create a dead band: small changes in sensor current may not produce a reliably observable output change until the offset is overcome. Offset can also contribute to output shifts associated with the sensor’s internal resistance. Low input bias current helps reduce measurement error when sensing low current levels.

Designing the feedback network

Set gain without exceeding the output range

First, select RF so the maximum expected sensor current does not push the amplifier beyond its linear output range. In the example, the sensor current spans 50 nA to 25 µA, and the supply is 3.3 V. The stated output limit for the LPV821 is 0.1 V below the positive supply, or 3.2 V. A 127 kΩ feedback resistor produces approximately 3.18 V at 25 µA, keeping the output within that limit. This check is essential: if the output saturates, the voltage no longer represents changes in sensor current as intended.

Check stability before choosing the final capacitor

Next, determine the minimum CF required for circuit stability, taking into account RF, the sensor capacitance, and the amplifier’s gain-bandwidth product. For the example, the calculation uses a 127 kΩ feedback resistor, 1 mF sensor capacitance, and the LPV821’s 8 kHz gain-bandwidth product. The minimum stable capacitance is a design constraint; the capacitor should not be reduced below it merely to increase bandwidth.

Verify the low-pass bandwidth

Finally, check that the selected RF and CF provide the required closed-loop bandwidth. With RF at 127 kΩ and CF at 1 µF, the feedback network’s low-pass pole is approximately 1.25 Hz—above the 1 Hz design target. If the selected capacitor produces a pole below the target, reduce CF only as far as the stability requirement permits. If the minimum stable CF still results in a bandwidth that is too low, the design may require a lower RF or an amplifier with a higher gain-bandwidth product.

Practical design implications

This example illustrates why sensor interfacing requires more than converting current into voltage. Engineers need to account for the sensor’s current range, capacitance, load requirements, response time, and noise, as well as the amplifier’s output swing, offset, bias current, and bandwidth. A disciplined design sequence helps avoid common trade-offs:

  • Set RF to provide useful voltage gain while keeping the maximum-current output within the amplifier’s linear range.
  • Establish the minimum feedback capacitance needed for stability.
  • Confirm that the resulting low-pass bandwidth is suitable for the sensor’s response and system requirements.
  • Review offset and input bias current when measuring small sensor currents.

For gas-sensing systems, a stable, low-power TIA with appropriately limited bandwidth can condition a noisy, slow-changing sensor signal for downstream electronics. The practical task is to coordinate gain, stability, filtering, and measurement accuracy as one design problem. Checking each constraint against the actual sensor and amplifier requirements provides a sound basis for integrating the signal path into a monitoring or control system.