Measuring low liquid flow in narrow tubing can be important in semiconductor processes that use deionized water (D.I. water) and chemical solutions. The µLF-100 ultrasonic micro liquid flow meter is designed for this kind of application, with sensors for 3 mm and 6 mm outside-diameter PFA tubing. Its non-contact ultrasonic measurement approach, compact sensor assembly, and configurable outputs give engineers a way to monitor flow in small process lines without moving meter components.

How the µLF-100 measures liquid flow

The meter uses a time-difference calculation based on the propagation of ultrasonic waves. Rather than relying on moving parts, the instrument processes the ultrasonic signal digitally. The source specifications describe the measurement approach as ultrasonic wave propagation with digital signal processing (DSP).

With no moving parts in the meter, there are fewer mechanical components associated with routine maintenance. The manufacturer also identifies no recalibration requirement as a product feature. These characteristics may be relevant when selecting instrumentation for process lines where maintenance access and service planning matter.

The µLF-100 is intended for D.I. water and chemical solutions, and its sensors are specified for PFA tubing. The sensor assembly is molded for the selected line size, supporting installation on either 6 mm O.D. by 4 mm I.D. tubing or 3 mm O.D. by 2 mm I.D. tubing.

Ultrasonic micro liquid flow meter ranges and accuracy

Available measurement ranges depend on the sensor model. The µLF-100-6 is specified for 277 to 5,560 ml/min, while the µLF-100-3 is specified for 94 to 1,881 ml/min. The instrument lists accuracy as ±2% of reading within an operating range of 1.6 to 33 ft./sec. Repeatability is specified at ±0.2%.

These figures should be considered alongside the process conditions and installation arrangement. The published velocity range and the stated flow-rate range are separate specifications; when checking suitability, confirm the correct sensor size and compare the application requirements with the relevant limits.

The meter also provides a low-cutoff setting adjustable from 0 to 10% in 1% steps. This setting can be considered when configuring how the instrument handles low-flow conditions. The available analog output range is configurable within 0 to 33 ft./sec.

Outputs and operator display

The display/converter provides a two-line, 16-character LCD showing rate and total information. Units can be selected from options that include LPH, LPM, ml/min., m/s, cm/s, GPH, GPM, FPM, and FPS. Selecting units that align with plant conventions can make readings easier to interpret during operation and troubleshooting.

For integration with control or monitoring equipment, the µLF-100 provides a 4–20 mA analog output for instantaneous forward flow. The specified load resistance is 450 ohms or less. A separate open-collector, no-voltage contact output can be configured for accumulated forward-flow pulses or upper- and lower-flow-rate limits. Pulse scaling options include metric and U.S. customary units, from 1 ml through 1 L and from 0.001 G through 1 G.

These outputs allow the meter to provide both a continuous flow signal and selected pulse or limit information. Integrators should verify the intended signal use, scaling, and electrical interface against the receiving control-system requirements.

Installation and configuration considerations

The converter is designed for panel mounting and has an IP50 enclosure. The sensor is an in-line wetted assembly made from PFA tube, with a specified operating temperature of 10° to 70° C. The converter installation temperature range is 10° to 50° C. The listed power supply is 12–30 VDC, and power consumption is 3.6 watts or less.

For installation planning, the specified minimum straight runs are 10 pipe diameters upstream and 5 pipe diameters downstream. Use the applicable tube diameter when interpreting these distances, and account for the available space around the sensor and its connections. Sensor cabling is 5 m long, so the mounting location of the panel converter should be planned accordingly.

Setup is performed through software integrated into the display unit. The installer is guided through questions about pipe material and size, display configuration, and desired outputs. The molded sensor fixture is provided for the selected line size, helping make the mounting arrangement specific to the tubing used.

Applying the meter in a process design

Before specifying the instrument, engineers and system integrators should confirm several application details:

  • Liquid: Check that the process uses D.I. water or a chemical solution suitable for the stated application.
  • Tubing: Confirm PFA material and select the 3 mm or 6 mm outside-diameter sensor to match the line.
  • Flow: Compare the expected operating flow with the range listed for that sensor, and review the stated velocity range for the accuracy specification.
  • Installation: Allow for the required upstream and downstream straight runs, temperature limits, cable length, and panel space.
  • Controls integration: Define the required analog signal, contact-output function, pulse scaling, and display units.

The converter weighs approximately 1 lb., and the sensor approximately 0.5 lb. The supplied package includes the display/converter, sensors, and cabling.

Conclusion

The µLF-100 is a specialized option for measuring low liquid flow in small PFA process lines, particularly in semiconductor applications involving D.I. water and chemical solutions. Its two sensor sizes, digital ultrasonic measurement, selectable display units, and analog and contact outputs support different monitoring and integration needs. Careful matching of tubing, flow range, installation conditions, and control-system signals is essential to determine whether it fits a particular process.