High-frequency flexural ultrasonic transducers offer a way to extend air-coupled ultrasound beyond the frequencies commonly used by commercial flexural ultrasonic transducers (FUTs). This matters in industrial measurement because a higher operating frequency can provide a shorter ultrasonic pulse and wavelength, supporting improved measurement accuracy and sensitivity. Achieving useful output, however, requires more than simply driving a conventional transducer at a higher frequency: the plate and piezoelectric element must be designed as a coupled system.

A study of high-frequency flexural ultrasonic transducers (HiFFUTs) demonstrated operation at 102.1 kHz and examined how the piezoelectric disc’s diameter affects both transmission and reception. Its findings offer practical guidance for engineers and system integrators selecting or developing air-coupled ultrasonic sensors for ranging, flow measurement, and other applications.

How a flexural ultrasonic transducer works

A typical FUT consists of a piezoelectric ceramic disc bonded to an elastic plate, often the front face of a cap. Applying a time-varying voltage to the disc causes the plate to flex. In receiving mode, incoming ultrasound makes the plate vibrate, bending the piezoelectric element and generating a voltage.

The device operates at resonant modes of the combined plate-and-disc structure. A plate can have many modes, but the axisymmetric modes are commonly used because they provide useful vibration and radiation characteristics. In the HiFFUT design, the higher-frequency mode of interest was the plate’s (1,0) mode.

FUTs are valued for their relatively simple construction, low cost, and mechanical robustness. They can also operate with low voltage and power. Their common operating frequencies, however, are typically below 50 kHz, with 40 kHz a frequent nominal frequency. Designing for a higher frequency can create shorter pulses, which may help distinguish ultrasonic signals arriving at slightly different times.

Why high-frequency flexural ultrasonic transducers need careful design

Raising frequency also makes air-coupled measurement more challenging. Air attenuates higher-frequency ultrasound more strongly, while the transducer’s own response can reduce the amplitude of transmitted and received signals away from its nominal operating frequency. The design therefore needs to produce efficient vibration and maintain a useful signal at the receiver.

Classical plate theory provides a starting point for estimating a plate’s resonant frequencies and mode shapes. But the bonded piezoelectric disc changes the behaviour of the structure: the plate, adhesive, and disc vibrate together. Finite element analysis (FEA) can account for this interaction and help evaluate how design changes affect the complete transducer.

In the reported design process, the researchers used plate theory to guide the choice of a practical cap, then used FEA to study the influence of the piezoelectric disc diameter. The prototype used a titanium cap and a thin PZT-5H disc bonded to the inside of its flexing plate. The work focused on the higher-frequency (1,0) mode.

Piezoelectric disc diameter affects transmission and reception

One important result is that disc diameter does not affect transducer behaviour in a simple, linear way. The model showed a non-monotonic relationship between disc diameter and the resonant behaviour of the plate-disc assembly. It also showed that the diameter producing the greatest plate displacement during transmission is not necessarily the one that produces the strongest received signal in a pitch-catch system.

This distinction is important when designing an ultrasonic measurement system. A transmitter-only design can be optimized for its output, but a practical system must also account for how the receiving transducer responds and how the acoustic field travels between the two devices. In the modeled configuration, a disc diameter that maximized transmitter displacement did not maximize the voltage signal at the receiver.

The predicted received signal varied substantially as disc diameter changed. The analysis indicated that choosing an appropriate diameter for a pitch-catch pair could produce a signal amplitude around an order of magnitude greater than a non-optimized design. The specific optimum depends on the system configuration, so transmitter performance alone is not a sufficient design measure.

Prototype performance at 102.1 kHz

The prototype’s electrical impedance and vibration shape were measured to verify its resonant behaviour. Measurements identified resonances at 20.6 kHz and 102.1 kHz, corresponding to the (0,0) and (1,0) modes. These results were close to the FEA predictions.

In a comparison using a 5 V continuous sine wave, the HiFFUT produced an acoustic pressure of 8.4 Pa at 102.1 kHz. The 40 kHz commercial FUT used as a baseline produced 4.9 Pa at its operating mode. The measured -3 dB bandwidths were approximately 3.0 kHz for the HiFFUT and 0.9 kHz for the commercial FUT. These measurements characterize the devices under the reported test conditions; they should not be interpreted as a guarantee of performance in every installation.

The HiFFUT’s radiation pattern included a main lobe with a -3 dB beam angle of 19 degrees, along with secondary lobes. In a pitch-catch test with a 200 mm propagation distance, the HiFFUT pair operated at about 2.6 times the frequency of the commercial pair. The received HiFFUT signal was temporally shorter, which made a later echo distinguishable in the measurement where it was obscured by the longer commercial-FUT pulse.

Practical implications for industrial measurement

For engineers evaluating air-coupled ultrasound, the work highlights several useful design considerations:

  • Optimize the complete system. Assess transmitter displacement, receiver sensitivity, and the acoustic path together, particularly in pitch-catch arrangements.
  • Use modeling to guide prototypes. Plate theory helps estimate resonances, while FEA can account for the bonded piezoelectric disc and other components.
  • Match frequency to the application. Shorter pulses at higher frequencies may help separate signals, but increased attenuation in air remains a design challenge.
  • Validate with measurements. Impedance analysis, vibration mapping, and acoustic pressure measurements can help confirm that a prototype behaves as intended.

High-frequency flexural ultrasonic transducers can provide a practical route to air-coupled operation above the range typical of commercial FUTs. The central lesson is that frequency, transducer construction, and receiver response must be considered together. For industrial applications such as distance ranging and flow measurement, a systematic design approach—combining analytical estimates, FEA, and prototype characterization—can help balance signal strength, bandwidth, directivity, and pulse duration.