For engineers designing compact, battery-powered sensing systems, the performance of an optical sensor depends on more than its light source or detector alone. The geometry between those components can influence both the strength and the quality of the signal. A study of a reflectance-based organic PPG sensor illustrates this trade-off using a red organic light-emitting diode (OLED), a surrounding organic photodiode (OPD), and Bluetooth Low Energy (BLE) data transmission.
The prototype was designed to collect a photoplethysmogram (PPG) waveform from a fingertip while using low power. Comparing two OLED and OPD layouts highlighted a practical design consideration: reducing the distance between the emitter and detector improved signal-to-noise ratio (SNR) under low-power operation, although the relationship was not the same at higher drive currents.
How reflectance-based PPG sensing works
A PPG sensor shines light into tissue and detects some of the light that returns. As arterial blood volume changes with each heartbeat, the amount of light absorbed by the tissue changes, producing a varying component in the detected signal. The waveform also includes a relatively steady component associated with tissue, veins, capillaries, and other non-pulsatile structures.
In a reflectance arrangement, the light source and detector sit on the same side of the tissue. This differs from transmission sensing, where tissue is positioned between an emitter and detector. Reflectance can support placement on different body locations, while transmission requires the emitter and detector to be placed opposite one another, as at a fingertip or earlobe.
For the prototype, red OLED light peaked at 625 nm. The OPD’s external quantum efficiency peaked at about 37% at the same wavelength, providing spectral overlap between emitted and detected light. The design used one light source rather than multiple sources, with the aim of reducing the illumination load in a wireless pulse meter.
Organic PPG sensor geometry and optical trade-offs
Both test devices placed a circular OLED at the center of a ring-shaped OPD, fabricated together on a glass substrate. Each OPD had an area of 0.16 cm². Device 1 used an OLED area of 0.06 cm² and an OLED-to-OPD gap of 1.65 mm. Device 2 used a 0.03 cm² OLED and a 2 mm gap.
Optical modelling using ray tracing and Monte Carlo methods was used to estimate how light from each layout would travel through a simplified finger model and reach the detector. The simulation estimated that Device 1 would deliver more total optical power to the OPD than Device 2. These estimates informed the design comparison; the measured PPG performance was evaluated separately on a subject’s index finger.
The layouts differed in both OLED area and spacing, so the comparison should be understood as an evaluation of the combined designs, not a test of gap distance in isolation. Even so, the findings demonstrate why emitter-detector geometry is a useful design variable when developing a reflectance sensor.
Signal quality depends on operating conditions
Both devices produced a PPG waveform during testing. At an OLED driving current of 93.6 µA, the reported SNR was 46 dB for Device 1 and 45 dB for Device 2. At a lower current of 2.4 µA, the results separated more clearly: Device 1 produced a 0.7 mV peak-to-peak signal at 18 dB SNR, while Device 2 produced a 0.3 mV signal at 8 dB SNR.
The study reported that Device 1 achieved an 18 dB SNR at a power consumption of 8 µW. At the higher current, the shorter gap did not produce a higher SNR: the authors associated the result with increased DC noise on the OPD. This contrast matters for low-power sensor design. A geometry that benefits operation at a modest drive level may not provide the same advantage when the emitter is driven harder.
Other physical and operating factors also affect the received waveform. The sensor must be coupled to the skin to collect reflected light, but excessive pressure can compress blood vessels and weaken the pulsatile signal. The results therefore point to both optical layout and consistent sensor-to-skin contact as considerations when evaluating a wearable PPG assembly.
Designing the emitter and detector as a system
The OLED and OPD areas involve related trade-offs. Reducing OLED area can lower the area available to emit light, while increasing current density on a smaller emitter may affect its lifetime. Enlarging the OLED does not necessarily mean that all of its area contributes usefully to light returned from the tissue. Similarly, the OPD must be large enough and positioned appropriately to collect reflected light, but a larger detector area can increase DC noise.
Optical simulation can help assess these interactions before fabrication. For system integrators and sensor developers, the broader lesson is to evaluate the emitter, detector, spacing, drive current, and signal-conditioning chain together. The prototype’s electronics converted the OPD photocurrent to voltage with a transimpedance amplifier, filtered high-frequency noise and the DC component, amplified the AC PPG component, and digitized the signal.
Wireless transmission and practical implications
The selected device transmitted PPG waveforms to a PC host over BLE. The signal was sampled at 500 samples per second with 8-bit resolution. Although the BLE chip’s stated on-air data rate was 1,000 kbps, the measured maximum throughput between the module and receiving PC host was 256 kbps, at a minimum connection interval of 7.5 ms. These figures describe different parts of the communication path; the practical data rate to the host was lower than the chip’s on-air rate.
For battery-operated monitoring, the sensor’s optical and electronic power requirements must be considered alongside wireless operation and the quality of the data being sent. The comparison shows that component placement can help preserve useful signal quality at low emitter power, but the optimal arrangement depends on the operating point and noise characteristics.
Conclusion
This reflectance-based organic PPG sensor demonstrates a compact approach to wireless waveform monitoring using a red OLED and an organic photodiode. In the tested designs, the smaller emitter-to-detector gap was associated with better SNR at low drive current, including an 18 dB signal at 8 µW, while performance at higher current was influenced by DC noise. For engineers developing wearable or portable sensing systems, the practical takeaway is to optimize geometry and operating conditions together, validate signal quality at intended power levels, and account for contact pressure and end-to-end wireless throughput.