Dissolved oxygen sensors help process teams monitor oxygen where concentration can affect product quality, equipment integrity or process control. The Hach Orbisphere GA2400 and GA2800 EX are electrochemical sensors designed for dissolved oxygen measurement in ppm and ppb environments, with options for laboratory analysis of oxygen in liquid or gas phases. Their application range spans beverage production, water systems and demanding chemical environments.
Choosing between the models involves more than comparing the lowest detectable concentration. Engineers and system integrators also need to consider the process environment, wetted materials, pressure and temperature conditions, measurement range and maintenance requirements.
How the GA2400 and GA2800 EX fit process applications
The GA2400 is intended for a broad range of non-harsh applications. Examples include beer and soft-drink production, semiconductor-water rinsing and reactor coolant systems. The GA2800 EX is designed for harsher environments, including chemical, oil and petrochemical plants. It is available in stainless steel or Hastelloy, and GA2800 EX units carry ATEX certification.
Both models use electrochemical sensing and can be configured with different pre-mounted membrane kits. The selected kit determines the measurement range, allowing a setup suited to the oxygen levels expected in the process. Depending on the kit, the stated ranges extend from 0.1 ppb at the low end to 400 ppm at the high end. This range of options makes kit selection an important part of specifying the sensor.
Matching measurement range and accuracy to the process
Very low oxygen concentrations can matter in applications where operators need to control oxygen to help protect products or equipment and avoid corrosion. The GA2400 and GA2800 EX support low-level measurement, with a listed minimum range of 0.1 ppb for the 2956A-A membrane kit at 25 °C. Other kits cover different concentration or partial-pressure ranges, so the appropriate configuration depends on the required measurement window.
Accuracy also varies by kit and instrument configuration. The technical data defines accuracy as trueness under ISO 5725 and specifies it as the greater of ±1% of reading or a stated absolute value. For example, the 2956A-A range lists ±1% of reading or ±0.1 ppb for certain instruments, and ±1 ppb for others. These qualifications matter when evaluating performance near the low end of a range; the ±1% figure should not be treated as the sole accuracy value in every condition.
Response time is another consideration when selecting a membrane kit. At 25 °C, the listed time for a 90% signal change ranges from 7.2 seconds for the 2956A-A to 2.5 minutes for the 2935A-A. The 2952A-A is listed at 38 seconds, while the 29552A-A is listed at 90 seconds. These values help teams compare configurations, but should be considered alongside the required range and accuracy.
Cartridge-based maintenance reduces service time
Traditional membrane and electrolyte replacement processes can take more than 30 minutes. The GA2400 and GA2800 EX use a cartridge system containing the components needed for membrane and electrolyte solution change. The stated maintenance time for this procedure is less than five minutes.
For plant teams, a faster replacement procedure can simplify routine sensor service. Recharge kits are available with pre-filled cartridges and pre-mounted membranes for specified configurations. Selecting the correct kit is essential: the replacement membrane must match the measurement range and sensor setup required by the application.
Materials and installation conditions
The GA2400 uses stainless steel 1.4404 (AISI 316L) for wetted materials. The GA2800 EX is offered with Hastelloy or stainless steel. The product specifications state that no O-rings are in contact with the sample. This is a relevant detail when reviewing the sensor’s sample-contacting materials for a process.
Published operating limits and installation details should also be checked during specification. The data lists a temperature resistance of -5 to 100 °C and temperature compensation from -5 to 60 °C. Pressure resistance is listed as a minimum of 40 bar with the default PPS collar and 100 bar with a stainless-steel collar. Flow requirements depend on the flow chamber: specified recommendations range from 25 to 180 mL/min across the listed chamber configurations.
Practical selection checklist
Before specifying a GA2400 or GA2800 EX, engineering and operations teams can use the following questions to guide selection:
- What is the process environment? Consider whether the application is a non-harsh process or a harsh chemical, oil or petrochemical setting, and review the model’s certification and materials.
- What range is required? Match the membrane kit to the expected oxygen concentration or partial pressure, including the lower measurement limit.
- What accuracy and response time are needed? Check the kit-specific figures, instrument qualifications and measurement conditions rather than relying on a single headline value.
- What are the installation conditions? Confirm pressure, temperature and flow-chamber requirements against the process and installation design.
- How will maintenance be managed? Identify the correct cartridge or recharge kit and plan the replacement procedure for the selected membrane.
Conclusion
The GA2400 and GA2800 EX provide configurable electrochemical oxygen measurement for applications ranging from beverage and water processes to harsh industrial environments. Their cartridge-based membrane and electrolyte change is designed to take less than five minutes, while multiple membrane kits support different measurement ranges and response characteristics. A sound specification starts with the process conditions, then aligns model, materials, certification, membrane kit and installation requirements with the measurement task.