Characterizing a deep-UV LED emission pattern can be difficult with conventional imaging equipment. Standard CMOS and CCD cameras are generally poorly suited to direct deep-UV detection because their sensor structures and optical components absorb ultraviolet light. A fluorescence-based method offers an alternative: the LED illuminates a fluorescent target, which converts the UV excitation into visible emission that a standard camera can record.

This approach makes it possible to assess how intensity varies across a target and with angle, without requiring a UV-extended camera. It can help engineers and technical teams evaluate LED output for applications where illumination distribution matters, including optical microscopy and the development or characterization of LED packages and arrays.

Why measure a deep-UV LED emission pattern?

An LED emission pattern describes the intensity of light as a function of angle from the normal of the chip surface. The pattern depends in part on the LED’s packaging geometry. A planar package can produce a distribution that follows a Lambertian trend, in which intensity decreases with the cosine of the angle from the surface normal. Other package geometries can produce different distributions: hemispherical packages may be more isotropic, while parabolic packages can produce more directional output.

These differences have practical consequences. In microscopy, for example, an LED may illuminate a specimen directly without lenses, making the angular distribution relevant to illumination uniformity. Deep-UV LEDs also have potential uses in water purification and gas sensing, where understanding how their light is emitted can support evaluation of the source for a particular setup.

Predictive models can be useful, but a general model is difficult to apply across packaged LEDs with different geometries. Measurement can therefore provide information about the actual emitter under test. Direct camera measurement is challenging in the deep UV, while UV-sensitive back-thinned CCDs can address the sensitivity problem but may be costly.

How fluorescence-based measurement works

The method uses a fluorescent specimen as a wavelength converter. Deep-UV radiation excites the specimen, which then emits visible light. A camera records this visible fluorescence, and its intensity distribution serves as an indirect measurement of the LED’s UV illumination.

In the reported setup, a 280 nm LED illuminated a fluorescein-doped agarose block. The block was placed close to the LED, and an objective lens collected the fluorescence. Relay lenses focused the image onto a CMOS camera, while an emission filter in front of the camera blocked excitation light from reaching the image. A bandpass filter in front of the LED narrowed its electroluminescence spectrum.

Images were background-corrected, and intensity profiles were extracted across the field of view. The setup provided a 4.5 mm field of view. To measure angular output beyond this limited area, the LED was mounted on a rotating stage and turned about an axis through the chip. Fluorescence intensity was recorded at different angles and compared with a Lambertian distribution.

What the measurements showed

Across the measured field, fluorescence was brightest near the centre and reached about 60% of its maximum intensity at the edge. Changing the LED drive current affected the overall signal but did not change the measured intensity distribution. Increasing the distance between the LED and specimen could reduce variation across the field by narrowing the range of incident angles, but it also reduced the light reaching the specimen. The relationship between intensity and distance is an important consideration when balancing coverage against signal level.

Angular measurements showed maximum intensity at 0 degrees from the chip normal, falling to approximately half the maximum at 60 degrees. The data followed a Lambertian trend with a reported coefficient of determination of 99.6%. This result describes the tested LED and setup; it should not be treated as a universal pattern for deep-UV devices.

The method was also applied to a second LED emitting at 275 nm. Its measured output had a batwing-like pattern, with intensity peaking around plus and minus 20 degrees. The result agreed with the pattern described in the device datasheet, demonstrating that the technique could distinguish between the tested package types.

Measurement considerations and limitations

Because the camera images fluorescence rather than the UV source directly, the method depends on a key assumption: fluorescent material is distributed evenly throughout the specimen. Variations in the fluorescent target could otherwise be mistaken for variations in LED output. The resulting image also combines the effects of excitation and fluorescence imaging, rather than representing the LED emission alone.

Field of view and image sampling matter, too. The reported 4.5 mm field captured only part of the broader emission pattern. Rotating the LED provided angular measurements, while scanning the source across the field is another possible way to extend coverage. The original measurement setup also did not meet the Nyquist sampling criterion for the image detail under consideration. A separate higher-magnification configuration was used to measure the microscope’s point spread function and demonstrate deconvolution, but the resolution change would have been less than one pixel in the original emission-pattern setup.

These constraints mean that the technique is best understood as an indirect characterization method. Its usefulness depends on the target’s uniformity, the optical configuration, the region sampled and the measurement objective. For applications focused on a relatively small illumination field, the measured area may be relevant; broader characterization requires angular sampling or additional coverage.

Practical implications for LED evaluation

Fluorescence-based imaging provides a way to characterize deep-UV LED output with standard camera technology when direct UV detection is impractical. It can reveal field-level intensity variation and distinguish different angular distributions, including Lambertian and non-Lambertian patterns. The approach may also support evaluation of deep-UV illumination arrangements by placing suitable optics between the LED and fluorescent target.

For engineering teams, the central takeaway is to match the measurement method to the decision being made. A local field profile can help assess illumination variation over a specimen-sized area, while rotation or scanning is needed to examine a wider emission pattern. With attention to fluorescence uniformity, sampling and field-of-view limits, this method offers a practical route to comparing deep-UV LED output and packaging behavior.