Reliable oxygen measurement can be important wherever a process depends on controlling gas composition, maintaining an inert atmosphere, or monitoring oxygen at trace levels. The Oxytrans II oxygen transmitter is designed for both parts-per-million (ppm) and percentage measurement, with user-selectable ranges and a 4–20 mA output for integration into industrial control and monitoring systems. Its range options, sensor design, and installation details help determine whether it is suitable for a particular application.

Oxytrans II oxygen transmitter: measurement ranges and models

The series includes separate configurations for ppm and percentage oxygen measurement. The OxyT II-T model provides four selectable ranges: 0–10, 0–100, 0–1,000, and 0–10,000 ppm O2. This range selection allows users to match the instrument to the expected oxygen concentration rather than treating all trace measurements as the same operating condition.

Percentage models also offer multiple selectable ranges, grouped into three configurations:

  • OxyT II-P1: 0–0.25%, 0–2.5%, and 0–25% O2.
  • OxyT II-P2: 0–0.5%, 0–5%, and 0–50% O2.
  • OxyT II-P3: 0–1%, 0–10%, and 0–100% O2.

For engineers and integrators, selecting the model and range is a basic application decision. The expected oxygen concentration should fall within the chosen range, and the measurement resolution should be appropriate for the process. The listed resolution is 0.1 ppm for ranges below 100 ppm, 1 ppm for the 1,000 and 10,000 ppm ranges, and 0.1% for the 25% range.

Signal integration and operating characteristics

The transmitter provides a 4–20 mA DC signal and is described as a two-wire, loop-powered instrument. Its specified supply voltage is 10–35 VDC, with reverse-voltage protection up to 40 VDC. A typical load is 470 ohms. These details are useful when checking compatibility with a control-system input, power supply, and loop design.

HART communication is listed as the digital output. The instrument also has a six-digit alphanumeric display with a bar graph, providing a local indication alongside the transmitted signal. Alarm output values are specified as 3.8 mA standard or 23 mA if desired. Confirm the selected alarm configuration and the receiving system’s handling of those values as part of the integration process.

The device uses a micro-fuel cell sensor. The manufacturer lists reverse-voltage protection and temperature compensation among its features, as well as no cross sensitivity to hydrogen. The enclosure rating is IP65, and the stated operating temperature is 0°C to +50°C. Listed accuracy is ±2% of full scale at constant temperature and ±5% of full scale for the stated temperature condition. These specifications should be considered in the context of the selected range and the application’s operating environment.

Calibration and sample conditions

Calibration depends on whether the instrument is configured for ppm or percentage measurement. For ppm ranges, calibration uses a gas with a known oxygen concentration, typically nitrogen as the balance gas. Percentage ranges use air. The product information describes a one-point calibration intended to support startup, and also identifies zero calibration for measurement in the low-ppm range.

Sample handling is another practical consideration. The specified sample flow rate is 1–3 standard cubic feet per hour (scfh), with a maximum flow-through pressure of 10 bar. The process connection is a ¼-inch tube fitting. Maintaining appropriate sample conditions and using a calibration gas suited to the measurement range are important steps when planning installation and commissioning.

Applications and configuration choices

Listed applications include semiconductor manufacturing, gas production, environmental chambers, metallurgy, chemical processing, gloveboxes, heat treating, and bright annealing. Across these settings, the relevant measurement may be trace oxygen in an inert gas or oxygen concentration expressed as a percentage. Matching the transmitter’s range to the process is therefore central to evaluating its fit.

Glovebox versions are available across the listed ppm and percentage model families. The ordering information also identifies sensor options for inert-gas applications and acidic-gas applications where CO2 is present. The specified part numbers include 16T303 and 16P306 for inert-gas ppm and percentage ranges, respectively, and 16T304 and 16P307 for acidic-gas applications. More configurations may be available, so the intended gas composition and model should be confirmed during selection.

An optional explosion-protection version is also identified in the product information as Ex ia IIC T4 (ATEX). Suitability depends on the installation requirements and applicable site approvals; verify the exact configuration and certification documentation before specifying equipment for a hazardous area.

Applying oxygen measurement in a control system

For plant teams, a useful selection review covers the expected oxygen concentration, required resolution, gas composition, sample pressure and flow, and available loop power. Integrators should also account for the transmitter’s 4–20 mA signal, typical load, HART capability, alarm-value configuration, and the receiving system’s response to out-of-range or alarm conditions.

The Oxytrans II combines selectable ppm or percentage ranges with local indication and standard industrial signal interfaces. Its practical value depends on choosing the right model, calibration approach, sensor configuration, and installation conditions for the process. Reviewing those factors before commissioning helps ensure that oxygen readings are meaningful to operators and usable by the control system.